119 participants who were attending electroencephalographic (EEG) videotelemetry as part of their clinical management of their epilepsy consented to take part in the trial. 683 epileptic seizures (77 clinically significant seizures) and 2648 normal physiological events were captured. When using pre-specified cut-off point 25 % heart rate change and/or oxygen desaturation 85 % on the basis of one/other, the device showed a sensitivity of 87 % for detecting clinically significant seizures. False Alarm Rate 4.5 (24 h FAR), d