2 Nm vs 21.8 Nm and 38.6 Nm,P .01), which led to more comminution. The CPT- and C-Stem-implanted femurs, although fracturing earlier, fractured in a simple pattern with less comminution. The differences in stem design mean higher stress at the critical point of failure in the CPT implanted femur compared with the Exeter and DePuy, which is likely the reason behind the observed increased risk of PPFs with the CPT implant. The CPT- and C-Stem-implanted femurs, although fracturing earlier, fractured in a simple pattern with less co