Translate   7 w

https://www.selleckchem.com/products/kya1797k.html
d SI joint motion. S2AI fixation decreased the S1 screw bending moments compared with fixation ending at S1. These benefits were paired with increased rod strain at L5-S1. Supplementation of S2AI fixation with SIJF implants provided further reductions (approximately 30%) in the sagittal plane and lateral bending SI joint motion compared with fixation ending at the S2AI position. This stability was not paired with significant changes in rod or screw strains. Constructs ending with an S2AI screw versus an S1 screw tended to be more stabl

  • Like
  • Love
  • HaHa
  • WoW
  • Sad
  • Angry