Translate   12 w

https://www.selleckchem.com/pr....oducts/Acadesine.htm
To investigate the suitability of the microDiamond detector (mDD) type 60019 (PTW-Freiburg, Germany) to measure the anisotropy function F(r,θ) of High Dose Rate (HDR) Ir brachytherapy sources. The HDR Ir brachytherapy source, model mHDR-v2r (Elekta AB, Sweden), was placed inside a water tank within a 4F plastic needle. Four mDDs (mDD1, mDD2, mDD3, and mDD4) were investigated. Each mDD was placed laterally with respect to the source, and measurements were performed at radial distances r=1cm, 3 and 5cm, and polar angles θ from 0° to 168

  • Like
  • Love
  • HaHa
  • WoW
  • Sad
  • Angry