Translate   4 w

https://www.selleckchem.com/
The purpose of this work was to assess technical performance of a prototype high-ratio (r29), 80 line cm-1grid for imaging conditions which mimic those for adult cardiovascular angiography. The standard equipment r15, 80 line cm-1grid was used as a reference. Plastic Water®LR phantoms with thickness in the range 20-44 cm were used to simulate adult patient attenuation and scatter. Grids were tested using x-ray field of view 20 and 25 cm and x-ray source to detector distance (SID) 107 and 120 cm. The primary transmission fraction (TP) was measured using both

  • Like
  • Love
  • HaHa
  • WoW
  • Sad
  • Angry